Setting Reference & Tolerance Values

When you assign a test list to a unit each of the tests from that list are assigned a record of the reference value and tolerance/action levels on that unit. Reference values and tolerance/action levels are unit specific so you must set these values for every unit a test is assigned to. Initially the reference value and tolerance/action levels are set to null so you will need to update them if you want to subject your tests to pass/fail criteria when performing a test list.

To set the reference and tolerance values click on the Set References and Tolerances link from the main admin page. From here you will see a list of all the test/unit pairs for any test list which has been assigned to a unit. To set the reference value for a given test on a unit you can either browse through the list or use the search box at the top of the page to find the test you are looking for. Note, you may also use the filters at the side of the page to narrow your search. The image below shows a search for a test which has been assigned to two units.

Searching for a test on a unit

Searching for a test on a unit

Once you have located the test/unit pair you want to set the reference for, click on its name. From here you will see information about the current reference value and tolerance/action levels including when these values were last changed and who changed them.

Reference and tolerance example

Reference and tolerance example

Numerical References & Tolerances

If the test you are setting reference values for is numerical (e.g. simple numerical or composite test types) you can choose a tolerance/action level from the Tolerances: drop down and set the reference value in the text box below that.

The tolerance/action levels of a test act as a window around the reference value. For example if a certain test on a unit has reference value of 50 and a tolerance/action level of +/- 2%/3% (i.e. lower action = -3%, lower tolerance = -2 %, upper tolerance = 2%, upper action = 3%) then values between 49 and 51 will be considered within tolerance, values between 48.5 & 49 or 51 & 51.5 will be considered at tolerance and values less than 48.5 or greater than 51.5 will be considered failing.

Boolean References

For boolean tests you choose whether the reference value is Yes or No and whether a mismatch between the reference value and the users selection should result in a warning or a failure.

Reference and tolerance example for boolean test

Reference and tolerance example for boolean test

Multiple Choice References

Multiple choice tests have no reference value. For information about multiple choice tolerances, please see the tolerances page.

History

As of version 0.3.0 users may add a comment when updating references and tolerances and this comment will be shown with the history of changes for the reference and tolerance.

Reference and tolerance history

Reference and tolerance history

Setting multiple References & Tolerances at the same time

To set a group of tests to the same reference/tolerance value, go to the Set References & Tolerances page and select all the tests you want to set:

Setting multiple references & Tolerance

Setting multiple references & Tolerance

then select Set multiple references and tolerances from the Actions dropdown list at the top and click Go.

On the next page set the desired reference & tolerance value then click Set tolerances and references.

Setting multiple references & Tolerance

Setting multiple references & Tolerance

Copying References & Tolerances between units

It is possible to copy references and tolerances from one unit to another. This is quite handy when you have the same test list assigned to multiple units with the same reference values set.

From the main admin page click on the Copy References & Tolerances link in the QC section.

Copy ref tols link

Copy ref tols link

On the next page, choose the source unit and test list (cycle) and the destination unit and click Copy references and tolerances.

Copy ref tols select

Copy ref tols select

Finally, you will be asked to review and confirm before the new references are set.

Copy ref tols confirm

Copy ref tols confirm